The MEDITT medical gas outlet is designed to meet NFPA-99 requirement
All outlets shall be tested and cleaned for oxygen service
Latch-valve assembly front panel rectangular assembly with primary check valve
The rough-in assembly shall contain a double seal to prevent gas leakage between the rough-in and latch-valve assembly after the wall is finished
The latch-valve assembly shall be indexed to the corresponding rough-in assembly to avoid accidental cross-fitting
3/8" (or 1/2") outside diameter type K copper inlet tube tube to be swiveled 360° for attachment to the piping system
A die-cast, light gray, epoxy powder-coated cover plate shall be provided to trim each wall / ceiling outlet and to fill the space between adjacent outlets. Cover plate design shall allow latch valves to be individually removed for servicing multiple exposed outlet may be built into a single box according to customer specification gas specific indexing
Good design practices should include a slide for each vacuum outlet
US colors available : Oxygen / Green Nitrous Oxide / Blue Medical Air / Yellow Vacuum / White Nitrogen / Gray Carbon Dioxide / Black Evacuation / Purple
Outlets available : OHMEDA type、JIS type、PURITAN BENNETT type、MEDITT type and DISS type
Medical Gas Outlet exposed type
The MEDITT medical gas outlet is designed to meet NFPA-99 requirement
Multiple exposed outlet may be built into a single box according to customer specification
3/8” (or 1/2”) outside diameter type K copper inlet tube to be swiveled 360° for attachment to the piping system
Gas specific indexing
Square patient consoles is equipped with built-in gas outlet, electric receptacle or nurse call etc.
The system encompassing extruded aluminum ducts with separate cavities
The exposed type console can serve single gas outlet or more (optional)
The exposed type console is designed for easy installation
US colors available : Oxygen / Green Nitrous Oxide / Blue Medical Air / Yellow Vacuum / White / Nitrogen / Gray Carbon Dioxide / Black Evacuation / Purple
Outlets available : OHMEDA type JIS type PURITAN BENNETT type MEDITT type DISS type
MEDICAL GAS BED-HEAD UNIT TYPE
The MEDITT medical gas outlet is designed to meet NFPA-99 requirement
Multiple exposed outlet may be built into a single box according to customer specification
3/8" (or 1/2")outside diameter type K copper inlet tube to be swiveled 360° for attachment to the piping system
Gas specific indexing
Square patient consoles is equipped with built-in gas outlet, electric receptacle or nurse call etc
The system encompassing extruded aluminum ducts with separate cavities
The exposed type console can serve single gas outlet or more (optional)
The exposed type console is designed for easy installation
The bed head unit is a system encompassing extruded aluminum ducting with 4 separate cavities the bed head unit is designed for oxygen air vacuum medical gas and communication cables
The bed head may include night light up and read light power with optional electric receptacle and around socket or nurse call telephone
The bed head is designed for easy installation compensate for irregularities of the wall on which it is installed
The bed head is color printed or anodized decorative strip from MEDITT selection or by special request
The behead unit can serve up to 2,3 and 4 beds at once with up to 4 gas outlets for each bed
Us colors is available : Oxygen / Green Nitrous Oxide / Blue Medical Air / Yellow Vacuum / White Nitrogen / Gray Carbon Dioxide / Black Evacuation / Purple
Outlets are available : OHMEDA type JIS type PURITAN BENNETT type MEDITT type DISS type
Half-Length General Care Modular Wall Outlet
The modular wall is ideal for ward、therapeutics room etc
Combinations of medical gas outlet、power receptacle and nurse call provision are optional with customer
Width 622mm,height is optional with customer
General Care Modular Wall Outlet
The modular wall is ideal for emergency room、intensive care unit、recovery room etc
Combinations of medical gas outlet、power receptacle and nurse call provision are optional with customer
Equipped with movable slide for monitor shelf、I.V. pole and storage basket etc
Height and width are optional with customer
Head Wall Outlet
The modular wall is ideal for hemodialysis room、neonatal Intensive Care Unit etc
Combinations of medical gas outlet、power receptacle and nurse call provision are optional with customer
Equipped with horizontal and vertical gliding slide for monitor shelf or storage basket
Hose Reel Gas Outlet
Combinations of medical gas outlet、length of hose are optional with customer
Includes rough-in box for immediate installation
Medical Gas Alarm
Master Alarm
The panel is equipped with on / off lamp、high pressure lamp、normal pressure lamp、low pressure lamp 、test button、silence button and buzzer .
When an alarm condition is created, both audible and visual alarm is actuated .
“TEST”button tests if all function is in working order .
“SILENCE”button can quiet the audible alarm and reset the alarm .
Without pressure gauge .
Represent the source equipment pressure .
Can be fitted with any number of master alarms .
Each alarm monitors one gas .
Gas service:oxygen、nitrous oxide、medical air、vacuum and nitrogen, each alarm is clearly marked and identified with letter and color for the gas specified .
The box material is aluminum .
Power 110 V / 220 V , 50/60 Hz .
Power consumption 3~5 Watt .
The audible alarm is rated for 85 dB at one meter away .